您好,欢迎来到宝玛科技网。
搜索
您的当前位置:首页OIML-R076-2-e07

OIML-R076-2-e07

来源:宝玛科技网
)E( 7002 noitidE 2-67R LMIOINTERNATIONALOIMLR76-2RECOMMENDATIONEdition 2007 (E)Non-automatic weighing instrumentsPart 2: Test report format

Instruments de pesage à fonctionnement non automatiquePartie 2: Format du rapport d'essai

ORGANISATIONINTERNATIONALEDEMÉTROLOGIELÉGALEINTERNATIONALORGANIZATIONOFLEGALMETROLOGYOIML R 76-2: 2007 (E)

Contents

Foreword ................................................................................................................................................................3 Introduction.............................................................................................................................................................4 Explanatory notes....................................................................................................................................................5 General information concerning the type................................................................................................................6 Information concerning the test equipment used for type evaluation......................................................................8 Summary of type evaluation....................................................................................................................................9 1 Weighing performance............................................................................................................................10 2

Temperature effect on no-load indication................................................................................................11

3 Eccentricity..............................................................................................................................................12 3.1 Eccentricity using weights.......................................................................................................................12 3.2 Eccentricity using a rolling load..............................................................................................................13 4 Discrimination and sensitivity.................................................................................................................14 4.1 Discrimination.........................................................................................................................................14 4.2 Sensitivity (non-self-indicating instrument)............................................................................................15 5 Repeatability............................................................................................................................................16 6 Time-dependence.....................................................................................................................................17 6.1 Zero return...............................................................................................................................................17 6.2 Creep.......................................................................................................................................................18 7 Stability of equilibrium............................................................................................................................19 8 Tilting......................................................................................................................................................20 9 Tare (weighing test).................................................................................................................................21 10 Warm-up time..........................................................................................................................................22 11 Voltage variations....................................................................................................................................23 12 Electrical disturbances.............................................................................................................................24 12.1 AC mains voltage dips and short interruptions........................................................................................24 12.2 Electrical bursts.......................................................................................................................................25 12.3 Surges......................................................................................................................................................27 12.4 Electrostatic discharges...........................................................................................................................29 12.5 Immunity to radiated electromagnetic fields...........................................................................................32 12.6 Immunity to conducted radio-frequency fields........................................................................................34 12.7 Disturbance tests for instruments powered from a road vehicle power supply........................................35 13 Damp heat, steady state...........................................................................................................................37 14 Span stability...........................................................................................................................................40 15 Endurance................................................................................................................................................46 16

Examination of the construction of the instrument..................................................................................48

17 Checklist..................................................................................................................................................49 17.1 All types of weighing instruments except non-self-indicating instruments.............................................50 17.2 Instruments for direct sales to the public and price computing and labeling instruments.......................57 17.3 Electronic weighing instruments.............................................................................................................60 17.4 Software-controlled digital devices and instruments...............................................................................61

2

OIML R 76-2: 2007 (E)

Foreword

The International Organization of Legal Metrology (OIML) is a worldwide, intergovernmental organization whose primary aim is to harmonize the regulations and metrological controls applied by the national metrological services, or related organizations, of its Member States. The main categories of OIML publications are:

󰂃 International Recommendations (OIML R), which are model regulations that establish the

metrological characteristics required of certain measuring instruments and which specify methods and equipment for checking their conformity. OIML Member States shall implement these Recommendations to the greatest possible extent; 󰂃 International Documents (OIML D), which are informative in nature and which are intended to

harmonize and improve work in the field of legal metrology; 󰂃 International Guides (OIML G), which are also informative in nature and which are intended to

give guidelines for the application of certain requirements to legal metrology; and 󰂃 International Basic Publications (OIML B), which define the operating rules of the various OIML

structures and systems. OIML Draft Recommendations, Documents and Guides are developed by Technical Committees or Subcommittees which comprise representatives from the Member States. Certain international and regional institutions also participate on a consultation basis. Cooperative agreements have been established between the OIML and certain institutions, such as ISO and the IEC, with the objective of avoiding contradictory requirements. Consequently, manufacturers and users of measuring instruments, test laboratories, etc. may simultaneously apply OIML publications and those of other institutions.

International Recommendations, Documents, Guides and Basic Publications are published in English (E) and translated into French (F) and are subject to periodic revision.

Additionally, the OIML publishes or participates in the publication of Vocabularies (OIML V) and periodically commissions legal metrology experts to write Expert Reports (OIML E). Expert Reports are intended to provide information and advice, and are written solely from the viewpoint of their author, without the involvement of a Technical Committee or Subcommittee, nor that of the International Committee of Legal Metrology. Thus, they do not necessarily represent the views of the OIML.

This publication – reference OIML R 76-2, Edition 2007 – was developed by Technical Subcommittee TC 9/SC 1 Non-automatic weighing instruments. It was approved for final publication by the International Committee of Legal Metrology in 2007 and supersedes the previous edition of R 76-2 (1993).

OIML Publications may be downloaded from the OIML web site in the form of PDF files. Additional information on OIML Publications may be obtained from the Organization’s headquarters: Bureau International de Métrologie Légale 11, rue Turgot – 75009 Paris – France Telephone: 33 (0)1 48 78 12 82 Fax: 33 (0)1 42 82 17 27 E-mail: biml@oiml.org Internet: www.oiml.org

3

OIML R 76-2: 2007 (E)

Introduction

The “Type evaluation report”, the subject of R 76-2, aims at presenting, in a standardized format, the results of the various tests to which a type of a non-automatic weighing instrument shall be submitted with a view to its approval. These tests are described in Annexes A and B of R 76-1.

All metrology services or laboratories evaluating types of non-automatic weighing instruments according to R 76-1 or to national or regional regulations based on OIML R 76-1 are strongly advised to use this “Type evaluation report”, directly or after translation into a language other than English or French. Its direct use in English or in French, or in both languages, is even more strongly recommended whenever test results may be transmitted by the country performing these tests to the approving authorities of another country, under bi- or multi-lateral co-operation agreements. In the framework of the OIML Certificate System for Measuring Instruments, and the OIML Mutual Acceptance Arrangement (MAA), use of this report format is mandatory, in French and/or in English with translation into the national languages of the countries issuing such certificates, if appropriate.

The “information concerning the test equipment used for type evaluation” shall cover all test equipment which has been used in determining the test results given in a report. The information may be a short list containing only essential data (name, type, reference number for purpose of traceability). For example:

󰂃 󰂃 󰂃 󰂃 󰂃

Verification standards (accuracy, or accuracy class, and no.);

Simulator for testing of modules (name, type, traceability and no.); Climatic test and static temperature chamber (name, type and no.); Electrical tests, bursts (name of the instrument, type and no.);

Description of the procedure of field calibration for the test of immunity to radiated electromagnetic fields.

Note concerning the numbering of the following pages

In addition to a sequential numbering: “R 76-2 page .....” at the bottom of the pages of this publication, a special place is left at the top of each page (starting with the following page) for numbering the pages of reports established following this model; in particular, some tests (e.g. weighing performance) shall be repeated several times, each test being reported individually on a separate page following the relevant format; in the same way, a multiple range instrument shall be tested separately for each range and a separate form (including the general information form) shall be filled out for each range. For a given report, it is advisable to complete the sequential numbering of each page by the indication of the total number of pages of the report.

4

OIML R 76-2: 2007 (E) Report page …./….

Non-automatic weighing instruments Type evaluation report EXPLANATORY NOTES =Meaning of symbols: I Indication nth indication In L Load ΔL = Additional load to next changeover point P I + ½ e – ΔL = Indication prior to rounding (digital indication) E I – L or = P – L or = I + ½ e – ΔL – L = Error = Corrected error Ec mpe = Maximum permissible error (absolute value) EUT = Equipment under test The name(s) or symbol(s) of the unit(s) used to express test results shall be specified in each form. PASSED X – For each test, the “SUMMARY OF TYPE EVALUATION” and the =“CHECKLIST” shall be completed according to this example: ===when the instrument has passed the test: =when the instrument has failed the test: when the test is not applicable: FAILED X – The white spaces in boxes in the headings of the report should always be filled in according to the following example: At start At max At end 21.2 °C Temp.: 20.5 Rel. h.: % Time: Bar. pres.: hPa where: Temp. temperature Rel. h. = relative humidity Bar. pres. = barometric pressure (barometric pressure is necessary for the span stability test and when specified by IEC test provisions; in other cases it may be necessary only for class I instruments). “Date” in the test report refers to the date on which the test was performed. In the disturbance tests (12.1 through 12.7), faults greater than e are acceptable provided that they are detected and acted upon, or that they result from circumstances such that these faults shall not be considered as significant (see T.5.5.6 in R 76-1); an appropriate explanation shall be given in the column “Yes (remarks)”. Numbers in brackets refer to the corresponding subclauses of OIML R 76-1. 5

OIML R 76-2: 2007 (E) Report page …./….

General information concerning the type Application no.: ..…………………..………………..……………… Type designation: ..…………………..………………..……………… Manufacturer: ..…………………..………………..……………… Applicant: ..…………………..………………..……………… Instrument category: ..…………………..………………..……………… error fraction pi = Complete instrument Module1 with Accuracy class2: Self-indicating Semi-self-indicating Non-self-indicating Min e = Max = d=n = =e = Max1 = d1=n1 = 1e2 = Max2 = d2=n2 = e3 = Max3 = d3=n3 = T = + T = – V Umin=VUmax=Vf = Hz Battery, Unom = V Unom = Zero-setting device: Tare device: Non-automatic Tare balancing Combined zero/tare device Semi-automatic Tare weighing Automatic zero-setting Preset tare device Initial zero-setting Subtractive tare Zero-tracking Additive tare Initial zero-setting range = % of Max Temperature range: °C Printer: Built-in Connected Not present No connection but connectable Instrument submitted: …………………….……… Load cell: …...………………………....... Identification no.: ………..…..………………. Manufacturer: …...………………………....... Software version: ……………………………. Type: …...………………………....... Connected equipment: ……………………………. Capacity: …...………………………....... ……………………………. Number: …...………………………....... Interfaces (number, nature): ……………………………. Classification symbol: …...………………………....... ……………………………. ……………………………. Remarks: …...………………………....... Evaluation period: ……………………………. …...………………………....... Date of report: ……………………………. …...………………………....... Observer: ……………………………. …...………………………....... 1 The test equipment (simulator or a part of a complete instrument) connected to the module shall be defined in the test form(s) used. 2 Please note that the class denominations used hereafter in this Recommendation do not include the oval around the number for improved clarity of the Test Report Format’s text. 6

OIML R 76-2: 2007 (E) Report page …./….

General information concerning the type

(continued)

Use this space to indicate additional remarks and/or information: Connection equipment, interfaces and load cells, choice of the manufacturer regarding protection against disturbances (5.1.1 a or 5.1.1 b of R 76-1), etc.

7

OIML R 76-2: 2007 (E) Report page …./….

Information concerning the test equipment

used for type evaluation

8

OIML R 76-2: 2007 (E) Report page …./….

Summary of type evaluation Application no.: Type designation: …………………………….....…………………………… ……………….…………………………………………… Report page PASSED FAILED Tests Remarks 1 Weighing performance Initial °C°C°C°C°C°C°C Temperature effect on no-load indication Eccentricity using weights Eccentricity using a rolling load Discrimination Sensitivity Repeatability Zero return Creep Printing, storage 7 Stability of equilibrium Zero-setting, tare balancing 8 Tilting 9 Tare 10 Warm-up time 11 Voltage variations 12.1 AC mains voltage dips and short interruptions a) Mains power supply lines 12.2 Electrical bursts b) I/O circuits and communication lines a) AC mains power supply 12.3 Surges b) Any other kind of power supply lines a) Direct application 12.4 Electrostatic discharges b) Indirect application (contact discharges only) 12.5 Immunity to radiated electromagnetic fields 12.6 Immunity to conducted radio-frequency fields a) Conduction along supply lines of external Electrical transients on 12 V and 24 V batteries 12.7 instruments powered from a b) Capacitive and inductive coupling via lines road vehicle power supply other than supply lines a) Initial test (at reference temperature) Damp heat, b) Test at high temperature and 85 % relative 13 steady state humidity c) Final test (at reference temperature) 14 Span stability a) Initial test 15 Endurance c) Final test EXAMINATIONS 16 Examination of the construction 17 Checklist 2 3.1 3.2 4.1 4.2 5 6.1 6.2 Remarks: 9

OIML R 76-2: 2007 (E) Report page …./….

1 WEIGHING PERFORMANCE (A.4.4) (A.5.3.1) (Calculation of the error) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (only class I) (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation Initial zero-setting > 20 % of Max: Yes E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Indication, I ΔL Load, L ↓ ↑ ↓ ↑ * No (see R 76-1, A.4.4.2) Error, E Corrected error, Ec ↓ * mpe ↑ ↓ ↑ Check if |Ec| ≤ |mpe| Passed Failed Remarks: 10

OIML R 76-2: 2007 (E) Report page …./….

2 Application no.: ………………………………………………………………………………….. Type designation: ………………………………………………………………………………….. Date: ………………………………………… Observer: ………………………………………… Verification scale interval, e: ………………………………………… Resolution during test (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation P = I + ½ e – ΔL Report Temp Zero indication,Add. load,Date Time *page (°C) I ΔL TEMPERATURE EFFECT ON NO-LOAD INDICATION (A.5.3.2) P ΔP ΔTemp. Zero-changeper … °C ΔP = difference of P for two consecutive tests at different temperatures ΔTemp. = difference of Temp. for two consecutive tests at different temperatures Check if the zero-change per 5 °C is smaller than e (class II, III or IIII) Check if the zero-change per 1 °C is smaller than e (class I) Passed Failed Remarks: * Give the report page of the relevant weighing test where weighing tests and temperature effect on no-load indication test are conducted together (see R 76-1, Figure 11). 11

OIML R 76-2: 2007 (E) Report page …./….

3 ECCENTRICITY (A.4.7) 3.1 Eccentricity using weights (A.4.7.1, 2 and 3) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Yes No 1) Test(s) performed on a mobile instrument (A.4.7.5): Yes No 2) In case of “Yes” to 1): A.4.7 and A.4.7.1 to A.4.7.4 have been applied: 3) In case of “No” to 2): Description of eccentricity test(s) (see A.4.7.5) under “Remarks” Location of test loads: mark on a sketch (see example below) the successive locations of test loads, using numbers which shall be repeated in the table below. 1 2 4 3 Also indicate in the sketch the location of the display or of another perceptible part of the instrument. Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* determined prior to each measurement Add. load, Indication, Error, Corrected Location Load, L I E error, Ec ΔL * * 1 *mpe * 2 * *3 * *4 Check if |Ec| ≤ |mpe| Passed Failed Remarks: 12

OIML R 76-2: 2007 (E) Report page …./….

3.2 Eccentricity using a rolling load (A.4.7.4) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Number of sections of the divided load receptor Undivided load receptor Location of test loads for each section of the load receptor: mark on a sketch (see example below) the successive locations of test loads, using numbers which shall be repeated in the table below. Also indicate in the sketch the location of the display or of another perceptible part of the instrument. 1 2 3 Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Section Direction (Å / Æ) Location Load, L Check if |Ec| ≤ |mpe| Passed Failed Remarks: **** Indication, I Add. load, ΔL Error, E * Corrected error, Ec mpe * * * 13

OIML R 76-2: 2007 (E) Report page …./….

4 DISCRIMINATION AND SENSITIVITY 4.1 Discrimination 4.1.1 Digital indication (A.4.8.2) Application no.: …………………………………………………. Type designation: …………………………………………………. At start At max At end °C Date: …………………………………… Temp.: Observer: …………………………………..… Rel. h.: % Verification scale interval, e: ……………………………..……… Time: Scale interval, d: ……………………………..……… Bar. pres.: hPa Removed load Extra load, Indication, Indication, Add 1/10 d I2 – I1 Load, L = 1.4 d I2 I1 ΔL Check if I2 – I1 ≥ d Passed Failed Remarks: 4.1.2 Analog indication (A.4.8.1) Application no.: …………………………………………………. Type designation: …………………………………………………. At start At max At end °C Date: …………………………………… Temp.: Observer: …………………………………..… Rel. h.: % Verification scale interval, e: ……………………………..……… Time: Scale interval, d: ……………………………..……… Bar. pres.: hPa Extra load Indication, Indication, I2 – I1 Load, L I2 I1 = |mpe| Check if I2 – I1 ≥ 0.7 mpe Passed Failed Remarks: 14

OIML R 76-2: 2007 (E) Report page …./….

4.1.3 Non-self-indicating instrument (A.4.8.1) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Rel. h.: % ………………………………………… Time: ………………………………………… Bar. pres.: hPa Extra load, Indication, Visible Load, L I displacement* = 0.4 |mpe| * Mark a visible displacement by “+” Check if there is a visible displacement Passed Failed Remarks: 4.2 Sensitivity (non-self-indicating instrument) (A.4.9) Application No.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp: Rel. h: % ………………………………………… Time: ………………………………………… Bar. pres: hPa Extra load Permanent displacement of Load L indicating element = |mpe| mm Check if the permanent displacement is equal to or greater than: 1 mm for an instrument of accuracy class I or II 2 mm for an instrument of accuracy class III or IIII with Max ≤ 30 kg 5 mm for an instrument of accuracy class III or IIII with Max > 30 kg Passed Failed Remarks: mm mm 15

OIML R 76-2: 2007 (E) Report page …./….

5 REPEATABILITY (A.4.10) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Automatic zero-setting and zero-tracking device is: Non-existent In operation Load (weighing 1-10) Load (weighing 11-20) E= I + 1/2 e – ΔL – L Add. load, Add. load, Indication Indication E E of load, I of load, I ΔL ΔL 1 11 2 3 4 5 6 7 8 9 10 Emax – Emin (weighing 1-10) mpe Check if a) E ≤ mpe (3.6 of R 76-1) mpe Emax – Emin (weighing 11-20) 12 13 14 15 16 17 18 19 20 b) Emax – Emin ≤ absolute value of mpe (3.6.1 of R 76-1) Passed Failed Remarks: 16

OIML R 76-2: 2007 (E) Report page …./….

6 TIME-DEPENDENCE 6.1 Zero return (A.4.11.2) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp: Verification Rel. h: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres: hPa (only class I) (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range P = I + ½ e – ΔL Time of reading 0 min Load during 30 minutes = 30 min 35 min Check if Load, L0 Indication of zero, I0 Add. load, ΔL P0 = P30 = P35 = P Change after 30 minutes: |Δ(P30 – P0)| = Change 5 minutes later: |Δ(P35 – P30) | = For multiple range instruments keep instrument unloaded for further 5 minutes: a) |Δ(P30 – P0) | ≤ 0.5 e b) |Δ(P35 – P30) | ≤ e1 (for multiple range instruments only) Passed Failed Remarks: 17

OIML R 76-2: 2007 (E) Report page …./….

6.2 Creep (A.4.11.1) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp: Verification Rel. h: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres: hPa (smaller than e): ………………………………………… (only class I) P = I + ½ e – ΔL Add. load, Load, Indication, Time of reading P ΔP L I ΔL 0 min 5 min 15 min 30 min* 1 h 2 h 3 h 4 h ΔP = difference between P at the start (0 min) and P at a given time. * If condition a) is met, the test is terminated. If not, the test shall be continued for the next 3.5 hours and condition b) shall be met. Condition a): ΔP ≤ 0.5 e after 30 minutes; and ΔP ≤ 0.2 e between the indication obtained at 15 minutes and that at 30 minutes Condition b): ΔP ≤ absolute value of mpe during the period of 4 hours Check if condition a) or b) is fulfilled Passed Failed Remarks: 18

OIML R 76-2: 2007 (E) Report page …./….

7 STABILITY OF EQUILIBRIUM (A.4.12) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp: Verification Rel. h: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres: hPa (smaller than e): ………………………………………… (only class I) Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation In the case of printing or data storage: No. Load (about 50 % of Max) First printed or stored weight value after disturbance and command Reading during 5 s after print-out or storageminimum value maximum value 1 2 3 5 4 Check if the first printed or stored weight value does not deviate more than 1 e from the readings during 5 seconds after print-out or storage (only two adjacent values allowed) Passed Failed In the case of zero-setting or tare balancing: Zero-setting E0 = I0 + ½ e – ΔL – L0 **Indication, I0 Zero-load Load, L0 No.* Error, E0 Add. load, ΔL (10 e) (< 4 % of Max) after zero-setting 1 2 3 4 5 Tare balancing E0 = I0 + ½ e – ΔL – L0 Indication, I0, Tare load Load, L0 ** Error, E0 No.* Add. load, ΔL (10 e) after tare balancing (about 30 % of Max) 1 2 3 5 *4 Apply the zero or tare load, disturb the equilibrium and immediately release zero-setting or tare, apply L0 if necessary and calculate the error according to A.4.2.3/A.4.6.2 of R 76-1. Perform this five times. ** L0 (10 e) shall be applied only if an automatic zero-setting or zero-tracking device is in operation. L0 shall be applied after releasing tare or zero-setting, immediately after zero is displayed the first time. Check if E0 ≤ 0.25 e Passed Failed Remarks: 19

OIML R 76-2: 2007 (E) Report page …./….

8 TILTING (A.5.1, A.5.1.1-A.5.1.3) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Instrument with leveling device and level indicator Instrument with automatic tilt sensor Instrument without level indicator or automatic tilt sensor Mobile instrument with automatic tilt sensor Mobile instrument with Cardanic suspension Limiting value of tilting = Give (if appropriate on a separate sheet) a sketch of the load receptor showing the location of the level indicator or direction of the tilting, if provided. Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range Ev = Iv + ½ e – ΔLv – L (v = 1, 2, 3, 4, 5), Iv = Indication, ΔLv = additional load Ev 0 = error calculated at or near zero Ec v = Ev – Ev 0 with Load, L Iv = ΔLv = Ev0 = Iv = ΔLv = Ev = Ec v = (Max) Iv = ΔLv = Ev = Ec v = Reference position Tilted position with the limiting value of tilting 1 2 3 4 5 unloaded L = 2 e=|E1 0 – Ev 0|max = mpe =|Ec 1 – Ec v|max =mpe =|Ec 1 – Ec v|max =Check if the differences are a) ≤ 2 e for the unloaded instrument (not valid for class II instruments, if they are not used for direct sales to the public) b) ≤ absolute value of mpe for the loaded instrument Failed Passed Remarks: 20

OIML R 76-2: 2007 (E) Report page …./….

9 TARE (WEIGHING TEST) (A.4.6.1) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Automatic zero-setting and zero-tracking device is: E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Indication, I ΔL Load, L ↓ ↑ ↓ ↑ * *Non-existent Not in operation Out of working range In operation First tare load Error, E Corrected error, Ecmpe ↓ ↑ ↓ ↑ * * Second tare load Check if |Ec| ≤ |mpe| Passed Failed Remarks: 21

OIML R 76-2: 2007 (E) Report page …./….

10 WARM-UP TIME (A.5.2) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation Duration of disconnection before test: hours E = I + ½ e – ΔL – L E0 = error calculated prior to each measurement at or near zero (unloaded) EL = error calculated at load (loaded) Indication, Add. load, Time* Load, L I ΔL Error, E = EL – E0 mpe Unloaded 0 min Loaded Unloaded 5 min Loaded Unloaded 15 minLoaded Unloaded 30 minLoaded * counted from the moment an indication has first appeared. Check if |EL – E0| ≤ |mpe| Passed Failed Remarks: 22

OIML R 76-2: 2007 (E) Report page …./….

11 VOLTAGE VARIATIONS (A.5.4) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… (only class I) Mains power supply (AC), A.5.4.1 External or plug-in power supply device (AC or DC), A.5.4.2 Rechargeable battery power supply, (re)charge during the operation of the instrument is possible, A.5.4.2 Non-rechargeable and rechargeable battery power supply, (re)charge during the operation of the instrument is not possible, A.5.4.3 12 V or 24 V road vehicle battery power supply, A.5.4.4 V Umin=VUmax=Unom VCalculate lower and upper limits of applied voltages according to A.5.4. If a voltage-range (Umin / Umax) is marked, use = the average value as reference value. Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation Category of power supply (if an instrument has more than one power supply): ............................................................... E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero Add. load, U, Load, Indication, Error, Corrected mpe Voltage (V) L I E error, Ec ΔL 10 e = Reference value Lower limit Upper limit 10 e = 10 e = Category of power supply (if an instrument has more than one power supply): .................................................................. E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero Add. load, U, Load, Indication, Error, Corrected mpe Voltage (V) L I E error, Ec ΔL 10 e = Reference value Lower limit Upper limit Check if |Ec| ≤ |mpe| 10 e = 10 e = Passed Remarks: Failed 23

OIML R 76-2: 2007 (E) Report page …./….

12 ELECTRICAL DISTURBANCES 12.1 AC mains voltage dips and short interruptions (B.3.1) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… VUminVUmaxV Mains power supply voltage: Unom Power supply voltage for the test: Utest V Unom or the average value of Umin and Umax Disturbance Result Load Duration / Number ofRepetitionAmplitude number of disturbancesinterval (s)of Utest ≥ 10 ≥ 10 s cycles Without disturbance 0 % 0 % 40 % 70 % 80 % 0 % 0.5 1 10 25 250 250 = Indication, I Significant fault (> e) or detection and reaction No Yes (see remarks) Check if a significant fault occurred Passed Failed Remarks: 24

OIML R 76-2: 2007 (E) Report page …./….

12.2 Electrical bursts (B.3.2) a) Mains power supply lines Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Mains power supply voltage: Unom VUminVUmaxV Power supply voltage for the test: Utest V Unom or the average value of Umin and Umax Test voltage (bursts) on each connection of the mains power supply lines: 1 kV Duration of the test at connection and each polarity: 1 min Disturbance Result Bursts on connection Load L ↓ ground N ↓ ground PE Polarity Indication, I ↓ =ground Significant fault (> e) or detection and reaction No Yes (see remarks) Without disturbance positive X negative Without disturbance positive X negative Without disturbance positive X negative L = phase, N = neutral, PE = protective earth Check if a significant fault occurred Passed Failed Remarks: 25

OIML R 76-2: 2007 (E) Report page …./….

b) I/O circuits and communication lines Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Test voltage (bursts) on each cable/interface (I/O signals, data and control lines): 0.5 kV Duration of the test at each cable/interface and each polarity: 1 min Disturbance Result Load Bursts on cable/interface (Type, nature) Polarity/ disturbance Indication, I Significant fault (> e) or detection and reaction No Yes (see remarks) Without disturbance 1 positive negative Without disturbance 2 positive negative Without disturbance 3 positive negative Without disturbance 4 positive negative Without disturbance 5 positive negative Without disturbance 6 positive negative Without disturbance 7 positive negative Without disturbance 8 positive negative Without disturbance 9 positive negative Explain or make a sketch indicating where the clamp is located on the cable; if necessary, use additional page. Check if a significant fault occurred Passed Failed Remarks: 26

OIML R 76-2: 2007 (E) Report page …./….

12.3 Surges (B.3.3) a) AC mains power supply Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end Observer: ………………………………………… Temp.: °C Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): Surges on AC mains power supply lines Load Disturbance Result 3 positive and 3 negative surges Significant fault (> e) synchronously with AC supply voltage Indication, or detection and reaction Polarity angle I amplitude/ apply on 0° 90° 180° 270° NoYes (see remarks) 0.5 kV L ↓ N Without disturbance pos X neg pos X neg pos X neg pos X neg 1 kV L ↓ PE Without disturbance pos X neg pos X neg pos X neg pos X neg 1 kV N ↓ PE Without disturbance pos X neg pos X neg pos X neg pos X neg L = phase, N = neutral, PE = protective earth Check if a significant fault occurred Passed Remarks: Failed 27

OIML R 76-2: 2007 (E) Report page …./….

b) Any other kind of power supply Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Kind or type of power supply DC Other form Voltage Surges on other power supply lines 3 positive and 3 negative surges apply on L ↓ N L ↓ PE N ↓ PE amplitude Polarity Significant fault (> e) or detection and reaction No Yes (see remarks) Disturbance Result Load Indication, I Without disturbance 0.5 kV pos neg pos neg pos Without disturbance 1 kV Without disturbance 1 kV neg L = positive conductor , N = negative or neutral conductor, PE = protective earth Check if a significant fault occurred Passed Failed Remarks: 28

OIML R 76-2: 2007 (E) Report page …./….

12.4 Electrostatic discharges (B.3.4) a) Direct application Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Contact discharge Paint penetration Air discharges Discharges Result Significant fault (> e) Number of RepetitionTest Load Indication, or detection and reaction voltage Polarity discharges interval I (kV) ≥ 10 ≥ 10 s No Yes (remarks, test points) Without disturbance2 pos. 4 pos. 6 pos. 8 pos. (air discharges) Without disturbance2 neg. 4 neg. 6 neg. 8 neg. (air discharges) Check if a significant fault occurred Passed Failed Remarks: Note: If the EUT fails, the test point at which this occurs shall be recorded. 29

OIML R 76-2: 2007 (E) Report page …./….

b) Indirect application (contact discharges only) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Horizontal coupling plane Load Test voltage (kV) Discharges Result Significant fault (> e) Number of RepetitionIndication, or detection and reaction Polarity discharges interval I ≥ 10 ≥ 10 s No Yes (remarks, test points) Without disturbance 2 pos. 4 pos. 6 pos. Without disturbance2 neg. 4 neg. 6 neg. Vertical coupling plane Load Test voltage (kV) Discharges Result Significant fault (> e) Number of RepetitionIndication, or detection and reaction Polarity discharges interval I ≥ 10 ≥ 10 s No Yes (remarks, test points) Without disturbance 2 pos. 4 pos. 6 pos. Without disturbance2 neg. 4 neg. 6 neg. Check if a significant fault occurred Passed Failed Remarks:Note: If the EUT fails, the test point at which this occurs shall be recorded. 30

OIML R 76-2: 2007 (E) Report page …./….

Specification of test points of EUT (direct application), e.g. by photos or sketches

a) Direct application

Contact discharges:

Air discharges:

b) Indirect application

31

OIML R 76-2: 2007 (E) Report page …./….

12.5 Immunity to radiated electromagnetic fields (B.3.5) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Frequency range 26-2000 MHz if the test according to B.3.6 cannot be applied (no mains or I/O ports available) Frequency range 80-2000 MHz if the test according to B.3.6 is performed (see form no. 12.6) Rate of sweep: Material of load: Disturbance Result Load Frequency Antenna Polarizationrange (MHz) Facing EUT Indication, I Significant fault (> e) or detection and reaction No Yes (remarks) Without disturbanceFront Vertical Horizontal Right Left Rear Front Right Left Rear Front Vertical Horizontal Right Left Rear Front Right Left Rear Frequency range: 26-2000 MHz or 80-2000 MHz Field strength: 10 V/m Modulation: 80 % AM, 1 kHz, sine wave Note: If EUT fails, the frequency at which this occurs shall be recorded Check if a significant fault occurred Passed Failed Remarks: 32

OIML R 76-2: 2007 (E) Report page …./….

Description of the set-up of EUT, e.g. by photos or sketches:

33

OIML R 76-2: 2007 (E) Report page …./….

12.6 Immunity to conducted radio-frequency fields (B.3.6) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Rate of sweep: Load: Material of load: Result Frequency range Significant fault (> e) Cable / Interface Indication, (MHz) or detection and reaction I No Yes (remarks) without disturbance without disturbance without disturbance without disturbance without disturbance without disturbance without disturbance without disturbance without disturbance without disturbance Frequency range: 0.15-80 MHz RF amplitude (50 ohms): 10 V (e.m.f.) Modulation: 80 % AM, 1 kHz, sine wave Check if a significant fault occurred. Note: If the EUT fails, the frequency at which this occurs shall be recorded Passed Failed Remarks: 34

OIML R 76-2: 2007 (E) Report page …./….

12.7 Electrical transients on instruments powered from a road vehicle power supply (B.3.7) a) Conduction along supply lines of external 12 V and 24 V batteries Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… 12 V battery voltage 24 V battery voltage 12 V battery voltage Disturbance Result Load Test pulse Conducted voltage Indication, I Significant fault (> e) or detection and reaction No Yes (remarks) Without disturbance 2a 2b* 3a 3b 4 +50 V +10 V –150 V +100 V –7 V 24 V battery voltage Disturbance Result Load Test pulse Conducted voltage Indication, I Significant fault (> e) or detection and reaction No Yes (remarks) Without disturbance 2a 2b* 3a 3b 4 * +50 V +20 V –200 V +200 V –16 V Test pulse 2b is only applicable if the measuring instrument may be connected to the battery via the main (ignition) switch of the car, i.e. if the manufacturer of the measuring instrument has not specified that the instrument is to be connected directly (or by its own main switch) to the battery. Check if a significant fault occurred Passed Failed Remarks: 35

OIML R 76-2: 2007 (E) Report page …./….

Capacitive and inductive coupling via lines other than supply lines b) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… 12 V battery voltage 24 V battery voltage 12 V battery voltage Kind or type of other Disturbance Result lines Significant fault (> e) (no power supply Conducted Indication, or detection and reaction Load Test pulselines) voltage I No Yes (remarks) Without disturbance a –60 V b +40 V Without disturbance a –60 V b +40 V Without disturbance Kind or type of other lines (no power supply lines) a –60 V b +40 V 24 V battery voltage Disturbance Result Load Test pulseConducted voltage Indication, I Significant fault (> e) or detection and reaction No Yes (remarks) Without disturbance a –80 V b +80 V Without disturbance a –80 V b +80 V Without disturbance a –80 V b +80 V Check if a significant fault occurred. Note: If EUT fails, the frequency at which this occurs shall be recorded Passed Failed Remarks: 36

OIML R 76-2: 2007 (E) Report page …./….

13 DAMP HEAT, STEADY STATE (B.2) a) Initial test (at reference temperature) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Load, L Indication, I ΔL ↓ ↑ ↓ ↑ * Error, E ↓ * ↑ Corrected error, Ec ↓ ↑ mpe Check if |Ec| ≤ |mpe| Passed Failed Remarks: 37

OIML R 76-2: 2007 (E) Report page …./….

b) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Corrected mpe Load, L Indication, I Error, E error, Ec ΔL ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑ * (*) Test at high temperature and 85 % relative humidity Check if |Ec| ≤ |mpe| Passed Failed Remarks: 38

OIML R 76-2: 2007 (E) Report page …./….

c) Final test (at reference temperature) Application no.: …………………………………………………. Type designation: …………………………………………………. Date: ………………………………………… At start At max At end °C Observer: ………………………………………… Temp.: Verification Rel. h.: % scale interval, e: ………………………………………… Time: Resolution during test Bar. pres.: hPa (smaller than e): ………………………………………… Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Load, L Indication, I ΔL ↓ ↑ ↓ ↑ * Error, E ↓ * ↑ Corrected error, Ec ↓ ↑ mpe Check if |Ec| ≤ |mpe| Passed Failed Remarks: 39

OIML R 76-2: 2007 (E) Report page …./….

14 SPAN STABILITY (B.4) Application no.: ………………………………………………………………….. Type designation: ………………………………………………………………….. Verification scale interval, e: …………………… Resolution during test (smaller than e): …………………… Automatic zero-setting and zero-tracking device is: Zero load = Non-existent Not in operation Out of working range Test load = Automatic span adjustment device: Existent Non-existent At start At max Temp.: Rel. h.: Time: Bar. pres.: At end °C % hPa Measurement no. 1: Initial measurement Date: ………………………………………… Observer: ………………………………………… Location: ………………………………………… E0 = I0 + ½ e – ΔL0 – L0 Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load, Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. Average error = average (EL – E0) = (EL – E0)max – (EL – E0)min= 0.1 e= If |(EL – E0)max – (EL – E0)min| ≤ 0.1 e, the loading and reading will be sufficient for each of the subsequent measurements; if not, five loadings and readings shall be performed at each measurement. Remarks: 40

OIML R 76-2: 2007 (E) Report page …./….

Subsequent measurements Measurement no. 2: At start At max At end Date: ………………………………………… Temp.: °C Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Average error = average (EL – E0) = Remarks: Measurement no. 3: At start At max At end °C Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication E0 EL EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Remarks: Average error = average (EL – E0) = 41

OIML R 76-2: 2007 (E) Report page …./….

Subsequent measurements Measurement no. 4: At start At max At end °C Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Average error = average (EL – E0) = Remarks: Measurement no. 5: At start At max At end °C Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Remarks: Average error = average (EL – E0) = 42

OIML R 76-2: 2007 (E) Report page …./….

Subsequent measurements Measurement no. 6: At start At max At end °C Date: ………………………………………… Temp: Observer: ………………………………………… Rel. h: % Location: ………………………………………… Time: Bar. pres: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Average error = average (EL – E0) = Remarks: Measurement no. 7: At start At max At end °C Date: ………………………………………… Temp: Observer: ………………………………………… Rel. h: % Location: ………………………………………… Time: Bar. pres: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication EL E0 EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Remarks: Average error = average (EL – E0) = 43

OIML R 76-2: 2007 (E) Report page …./….

Subsequent measurements Measurement no. …. : At start At max At end Date: ………………………………………… Temp.: °C Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication E0 EL EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Average error = average (EL – E0) = Remarks: Measurement no. …. : At start At max At end °C Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Measurement after the temperature test Measurement after the damp heat test Measurement after disconnection from the mains Measurement after change in test location Other condition: ……………………………………………………………………………………….... Automatic span adjustment device activated (if existent) EL = IL + ½ e – ΔL – L Add. load, Add. load,Indication Corrected Indication E0 EL EL – E0 of load, IL value* of zero, I0 ΔL0 ΔL 1 2 3 4 E0 = I0 + ½ e – ΔL0 – L0 5 * When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks. If five loadings and readings have been performed: Remarks: Average error = average (EL – E0) = 44

OIML R 76-2: 2007 (E) Report page …./….

Measurement no. Plot on the diagram the indication of temperature test T , damp heat test D and disconnections from the mains power supply P …………………………………………………………………………………………………………… …………………………………………………………………………………………………………… 5 6 7 8 Maximum allowable variation 4 1 2 3 14 SPAN STABILITY (B.4) Type designation: Application no.: +1.5 e +1 e +0.5 e 0 –0.5 e –1 e –1.5 e Average error e 45

Passed Failed OIML R 76-2: 2007 (E) Report page …./….

15 ENDURANCE (A.6) Application no.: ………………………………………………………………………………….. Type designation: ………………………………………………………………………………….. Verification scale interval, e: ………………………………………… Resolution during test (smaller than e): ………………………………………… a) Initial test At start At max At end °C Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: % Location: ………………………………………… Time: Bar. pres.: hPa Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range In operation E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Add. load, Indication, I ΔL Load, L ↓ ↑ ↓ ↑ * Error, E ↓ * ↑ Corrected error, Ec ↓ ↑ mpe 46

OIML R 76-2: 2007 (E) Report page …./….

b) Performance of the test Number of loadings: Load applied: c) Final test At start At max Date: ………………………………………… Temp.: Observer: ………………………………………… Rel. h.: Location: ………………………………………… Time: Bar. pres.: Automatic zero-setting and zero-tracking device is: Non-existent Not in operation Out of working range E = I + ½ e – ΔL – L Ec = E – E0 with E0 = error calculated at or near zero* Durability error due to wear and tear = |Ec initial – Ec final|** Add. load, Indication, I ΔL Load, L ↓ * At end °C % hPa In operation Error, E ↓ * Corrected error, Ec ↓ ↑ mpe ↑ ↓ ↑ ↑ Durability error due to wear and tear** Check if the durability error due to wear and tear is ≤ mpe Passed Failed Remarks: 47

OIML R 76-2: 2007 (E) Report page …./….

16

Use this page to indicate any description or information pertaining to the instrument, additional to that already contained in this report and in the accompanying national type approval or OIML Certificate. This may include a picture of the complete instrument, a description of its main components, and any remark which could be useful for authorities responsible for the initial or subsequent verification of individual instruments built according to the type. It may also include references to the manufacturer.

Description:

Remarks:

EXAMINATION OF THE CONSTRUCTION OF THE INSTRUMENT

48

OIML R 76-2: 2007 (E) Report page …./….

CHECKLIST

This checklist has been developed based on the following principles:

󰂃

to include requirements that cannot be tested according to tests 1 through 15 above, but that shall be checked experimentally, e.g. the operating range of the tare device (4.6.4), or visually, e.g. the descriptive markings (7.1);

to include requirements which indicate prohibitions of some functions, e.g. automatic tare device for instruments for direct sales to the public (4.13.3.3);

to include neither general requirements, e.g. suitability for use (4.1.1.2), nor weights and verification devices, e.g. auxiliary verification devices (4.9);

not to include requirements that allow functions or devices to be used, e.g. a combined semi-automatic zero-setting and tare device operated by the same key (4.5.4).

󰂃 󰂃 󰂃

This checklist is intended to serve as a summary of the results of examinations to be performed and not as a procedure. The items on this checklist are provided to recall the requirements specified in R 76-1, and they shall not be considered as a substitution to these requirements.

As for non-self-indicating instruments, clause 6 of R 76-1 shall be followed in lieu of this checklist.

The requirements that are not included in this type evaluation report (tests 1-15 and checklist 17) are considered to be globally covered by the type approval or OIML Certificate (e.g. classification criteria [3.2 and 3.3], suitability for application, use and verification [4.1.1.1, 4.1.1.2 and 4.1.1.3]).

For non-mandatory devices, the checklist provides space to indicate whether or not the device exists and, if appropriate, its type. A cross in the box for “existent” indicates that the device exists and that it complies with the definition given in the terminology. When indicating that a device is non-existent, also check the boxes to indicate that the tests are not applicable (see page 5).

If appropriate, the results stated in this checklist may be supplemented by remarks given on additional pages.

49

OIML R 76-2: 2007 (E) Report page …./….

17 CHECKLIST Application no.: …………………………….…….………………………… Type designation: …………………………….…….………………………… 17.1 All types of weighing instruments except non-self-indicating instruments (6.1-6.9, R 76-1) PASSED Requirement procedures Testing FAILED Remarks Descriptive markings 7.1.1 A.3 Compulsory in all cases: manufacturer's mark or name accuracy class maximum capacity, Max, Max1, Max2,... (+3.3.1) minimum capacity, Min verification scale interval, e, e1, e2, ... (+3.3.1) 7.1.2 A.3 Compulsory if applicable: name or mark of manufacturer’s agent serial number identification marks on separate but associated units type approval mark scale interval, d (d < e) software identification (if applicable) maximum tare effect, T (subtractive tare only if T ≠ Max) maximum safe load, Lim (if Lim > Max + T) special temperature limits counting ratio ratio between weight platform and load platform range of plus/minus indication 7.1.3 A.3 Additional markings: not to be used for direct sales to the public to be used exclusively for: the stamp does not guarantee ..... / guarantees only ..... to be used only as follows: special applications clearly marked (weighings ranges in 3.2 classes I and II or II and III) near display “not to be used for direct sales to the public” (for 4.15 instruments similar to those used for direct sales to the public) 7.1.4 A.3 Presentation of markings: indelible easily readable grouped together in a clearly visible place Max, Min, e and d (if d ≠ e) on or near display permanently shown in a clearly visible position possible to seal and apply a control mark/removal will result in destruction markings B and G additional information shown alternatively on a plate or 7.1.4 and displayed by a software solution either permanently or 7.1.1 B, accessed by a simple manual command 7.1.2 G 7.1.5.1 A.3 Instruments with several load receptors and load measuring devices: identification mark, Max, Min and e of each load receptor on relating load measuring device (Lim and T = + if applicable) 50

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures Testing Remarks 7.1.5.2 A.3 Separately-built main parts: identification mark repeated in descriptive markings 4.1.1.3 Suitability for verification: identification of devices which have been subject to separate type examination Verification marks and sealing 7.2 A.3 Verification mark: cannot be removed easy application visibility without the instrument to be moved when it is in service 7.2.2 Verification mark support or space: which ensures conservation of the mark 2 for stamp, stamping area ≥ 150 mm for self-adhesive type, ø ≥ 15 mm 4.1.2.4 A.3 Securing of components and preset controls: location form 4.1.2.4 Securing with software means legal status of the instrument recognizable 4.1.2.4 a evidence of any intervention protection against changes of parameters and the reference 4.1.2.4 b numbers facilities for affixing the reference number 4.1.2.4 c 4.1.2.5 Span adjustment device (automatic or semi-automatic): Existent Non-existent external influence impossible after securing 4.1.2.6 Gravity compensation: Existent Non-existent external influence on or access to impossible after securing Documentation 8.2.1 A.1 Technical information and data: characteristics of the instrument 8.2.1.1, specifications of modules 3.10.2 fractions, pi (modules tested separately) 3.10.2.1 specifications of families 3.10.4 specifications of components applicable descriptive documents (according to nos. 1-11) 8.2.1.2 specific declaration of the manufacturer 5.3.6.1 A.1 limiting value of tilting defined by the manufacturer 3.9.1.1 8.2.2 A.2 Examination of: documents functions (spot checks) test reports from other authorities Indicating device 4.2.1 Reading: reliable, easy and unambiguous overall inaccuracy ≤ 0.2 e (analog indication) size, shape and clarity by simple juxtaposition 4.2.2.1 A.3 Units of: mass price 51

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures Testing Remarks 4.2.2.1 Form of indication: for one indication, one unit of mass kscale interval in the form (1, 2 or 5) × 10 same scale interval for all indicating devices, printing devices and tare weighing devices 4.2.2.2 Form of digital indication: at least one figure at right Decimal sign: shall maintain its position (scale interval changed automatically) separate at least one figure to the left and all to the right on one line with the bottom of the figures Zero: only one non-significant zero to the right for values with decimal sign, non-significant zero only in third position 4.2.3 Limits: preventing of indication above Max + 9 e preventing of indication below zero unless a tare device is in operation (–20 d is accepted) 4.2.4 “Approximate” displaying device: Existent scale interval > Max/100 without being smaller than 20 e 4.2.5 Semi-self indicating instruments: extension of self-indication range ≤ self-indication capacity Analog indication: thickness and length of scale marks 4.3.1 scale spacing 4.3.2 limit of movement below zero and above capacity of self- 4.3.3 indication damping of oscillations of displaying component 4.3.4 4.4.1 Changing of digital indication: after change in load, previous indication not longer than 1 s 4.4.2 Stable equilibrium of digital indication: printed or stored weight values do not deviate more than 1 e from the final weight value zero or tare operations are within their accuracy requirements no printing, data storage, zero-setting, or taring during continuous or temporary disturbance of equilibrium 4.4.3 Extended digital indication: Existent not allowed when there is a differentiated scale division displaying a smaller scale interval only during pressing a key at most, 5 s after manual command prevention of printing while the device is in operation 4.4.4 Digital indications other than primary indications: Existent additional indications do not lead to any ambiguity to primary indications quantities identified by units, symbols, signs or designations thereof weight values (not weighed) shall be clearly identified or display only temporarily on manual command and shall not be printed the inoperative weighing mode is clear and unambiguously recognizable 52

Non-existent Non-existent Non-existent

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures Testing Remarks 4.4.5 Digital printing: Existent Non-existent clear and permanent figures ≥ 2 mm high name or symbol of units to the right of the value above column of values printing impossible when equilibrium not stable 4.4.6 Memory storage: Existent Non-existent storage, transfer, totalizing, etc. inhibited when equilibrium not stable 3.4.1 Auxiliary indicating device (Classes I and II only; Existent Non-existent not allowed on multi-interval instruments) If existent, type: rider interpolation complementary differentiated scale division only to the right of decimal sign k3.4.2 d < e ≤ 10 d, e = 10 kg or e = 1 mg for class I with d < 1 mg Differences between results 3.6.3 Differences: between multiple indications: ≤ mpe between digital indications and printout: zero 3.6.4 between two results: ≤ mpe for same load when method of balancing changed (semi-self-indicating) Tilting of instrument of class II, III or IIII 3.9.1.1 a marking on the level indicator shows the limiting value of tilting level indicator fixed firmly in a place clearly visible to the user an automatic tilt sensor releases a display switch-off or other appropriate alarm signal and inhibits the printout and data transmission Zero-setting, -tracking and -indicating Existent Non-existentInitial zero-setting Automatic zero-setting Semi-automatic zero-setting Non automatic zero-setting Zero-tracking Zero-indicating Effect shall not alter Max 4.5.1 Overall effect of: zero-setting A.4.2.1 = % zero-tracking initial zero-setting = % 4.5.2 A.4.2.3 Accuracy: deviation ≤ 0.25 e 4.5.3 Multiple range: Existent Non-existent effective for greater weighing range (if switching when loaded possible) 4.5.4 Control of zero-setting: separate from that of tare weighing device Semi-automatic zero-setting: functions only in stable equilibrium and if it cancels any previous tare operation 53

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures 4.5.5 A.4.2.2 Testing Remarks Zero-indicating device (digital indication): shows deviation ≤ 0.25 e not mandatory if auxiliary indicating device or rate of zero- tracking ≥ 0.25 d/s 4.5.6 Automatic zero-setting: operates only when equilibrium stable and indication has remained stable below zero at least 5 seconds 4.5.7 Zero-tracking: operates only when indication at zero or at negative net value equivalent to gross zero and equilibrium stable corrections ≤ 0.5 d/s when operates after tare, the overall effect may be 4 % of Max Existent Non-existent Tare devices Tare weighing Tare balancing Combined zero-setting and tare balancing Tare indicating Type: Subtractive Additive applicable requirements from 4.1 through 4.4 are fulfilled 4.6.1 4.6.2 Tare weighing device: dT = d 4.6.3 A.4.6.2 Accuracy: ±0.25 e (electronic instruments and instruments with analog indication), e = e1 for multi-interval better than ±0.5 d (mechanical instruments with digital indication 4.6.4 Operating range: prevention of operation at its zero effect or below its zero effect prevention of operation above its maximum indicated 4.6.5 Visibility of operation: operation indicated net with sign “NET”, “Net”, “net” or complete word (digital indication) NET disappears if gross displayed temporarily tare value or letter “T” (mechanical additive tare device) 4.6.6 Subtractive tare: prevention of use above Max or indication that capacity is reached 4.6.7 Multiple range: operation effective in greater weighing ranges if switching when loaded possible tare values are rounded to the scale interval of the actual weighing range which is in operation 4.6.8 Semi-automatic or automatic tare: operation only in stable equilibrium 4.6.9 Combined zero/tare: accuracy (4.5.2) zero indicating device (4.5.5) zero-tracking (4.5.7) 54

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures 4.6.10 Testing Remarks Consecutive tare operations: indicated or printed tare weight values clearly designated (if tare devices operative at the same time) 4.6.11 Printing net or gross: without designation designation: by G or B (gross) by N (only net printed) designation of net and tare by N and T (if net printed with gross and/or tare) instead of G, B, N and T, complete words printing separately net and tare with identification (determined by different tare devices) Existent Non-existent Preset tare dT = d or automatically rounded to d 4.7.1 transferred from one range to another one with larger ei, shall be rounded to the latter (multiple range) tare value ≤ Max1 for the same net weight value (multi-interval) and calculated net value rounded to the scale interval for the same net weight value 4.6.10 applies 4.7.2 cannot be modified/cancelled if tare operated after the preset tare is still in use operates automatically if clearly identified with load 4.6.5 applies 4.7.3 possibility to indicate preset tare if calculated net printed then preset tare value is printed as well 4.6.11 applies designation of preset tare by PT or complete word Existent Non-existent Locking devices 4.8.1 Positions: only two stable positions weighing only in ‘weigh” position positions clearly shown 4.8.2 Existent Non-existent Multiple ranges 4.10 Selection of weighing ranges: range in operation clearly indicated selection from smaller to greater range possible at any load (manual) selection from smaller to the following greater range (automatic) possible only for load ≥ Maxi of smaller range selection from a greater to a smaller range (manually) or to the smallest range (automatically) only 󰂃 at no load when zero or negative net value is indicated 󰂃 tare is cancelled automatically 󰂃 zero is set to ±0.25 e1 automatically Selection between load receptors, transmitting and measuring devices Existent Non-existent compensation for unequal no-load effect 4.11, 4.11.1 4.11.2 zero-setting without ambiguity and in accordance with 4.5 4.11.3 weighing impossible while selection 4.11.4 combinations easy identifiable 55

OIML R 76-2: 2007 (E) Report page …./….

PASSED FAILED Requirement procedures Testing Remarks 4.12 “Plus and minus” comparator instruments 4.12.1 Distinction of zones: “+” and “–” signs (analog indication) by inscription (digital indication) 4.12.2 Scale: with at least one scale division, d = e on either side of zero and value of d = e shown at either end Mechanical counting instruments with unit weigh receptor 4.17.1 Scale: with at least one scale division, d = e on either side of zero and value of d = e shown on the scale 4.17.2 Counting ratio: shown clearly above each counting platform or each counting scale mark 4.20 Modes of operation: clearly identification of mode which is actually in operation manual switching back to weighing mode in any mode and at any time possible automatic selection of mode only within a weighing sequence automatic switching back to the weighing mode at the end of the weighing sequence zero indication after returning from switch-off condition automatic check of zero position before returning from switch-off condition 56

OIML R 76-2: 2007 (E) Report page …./….

17.2 Instruments for direct sales to the public and price computing and labeling instruments Requirement 4.5.4 Testing procedures PASSED FAILED 57

Remarks Miscellaneous checks (direct sales to the public) Combined semi-automatic zero-setting device and semi-automatic tare-balancing device operated by the same key: not allowed 4.8.1 “Preweigh” position: not allowed 4.13.10 Counting ratio: 1/10 or 1/100 (mechanical counting instrument) 4.13.5 Impossibility of weighing during: locking operation adding or subtracting weights 4.13.7 Auxiliary and extended indicating device: not allowed 1When significant fault has been detected (electronic instruments): 4.13.9 visible or audible alarm provided for customer and data transmission prevented until user takes action or cause disappears Indication device (direct sales to the public) 4.13.1, 4.13.6 Primary indications to both vendor and customer: two display sets, one vendor- and one customer display: Yes No one display set for vendor and customer Yes No weight information about correct zero position tare operation preset tare operation height of numerical figures displayed to the customer ≥ 9.5 mm Instruments to be used with weights: value of weights possible to distinguish Zero-setting device (direct sales to the public) 4.13.2 Non-automatic zero-setting: only allowed when operated with a tool Tare device (direct sales to the public) not allowed on mechanical instrument with weights receptor 4.13.3 on instruments with one platform public can see whether: - tare is in use - tare setting is altered only one tare shall be in operation at any given time while tare or preset tare is in operation recalling of gross values is prohibited 4.13.3.1 Non-automatic tare: displacement of 5 mm at most e 4.13.3.2 Semi-automatic tare: reduction of value of tare not permitted and canceling of tare effect only if no load on the receptor One of the following condition fulfilled: tare value indicated permanently in a separate display 1 Checked by verifying the compliance with documents or by simulating faults; this check does not duplicate the disturbance tests 12.1-12.7.

OIML R 76-2: 2007 (E) Report page …./….

Requirement Testing procedures PASSED FAILED 58

Remarks indicated with sign “–” when no load on the receptor tare effect cancelled automatically when unloading after net weighing 4.13.3.3 Automatic tare: not allowed 4.13.4 Preset tare: indicated on separate display clearly differentiated from weight display reduction of tare value not permitted and canceling of tare effect only if no load on the receptor impossible to operate if tare device in operation cancelled at the same time as PLU if associated with PLU Self-service instruments: with one set of scales or displays 4.13.11 two sets of scales or displays instrument has two sets of scales or displays Primary indications shall include the product designation if a ticket is printed Price computing instruments and price scales (direct sales to the public) Requirements of 4.13 for direct sales to the public are met 4.14 4.14.1 Supplementary primary indications (4.13.6) unit price price to pay if applicable number, unit price and price to pay for non- weighed articles, price totals 4.14.2 Price scales: 4.2 4.2 and 4.3.1-4.3.3 4.3.1-4.3.3 error of price scale |W × U – P| ≤ e × U 4.14.3 Price computing: multiplication of indicated weight and unit price as indicated rounding to the nearest interval of price to pay unit price: price/100 g or price/kg Indications of weights, unit price and price to pay visible: while load on load receptor and for at least 1 s after stable weight indication or after any introduction of unit price freezing for ≤ 3 s after removing load and not possible to introduce or change unit price (if indication has been stable before and would otherwise be zero) printing weight, unit price and price to pay Stored in memory: before printing same data not to be printed twice for customer 4.14.4 Additional functions for trade and management: all transactions are printed for customer they shall not lead to confusion 4.14.4.1 Prices-to-pay (positive or negative) of non-weighed articles: weight indication zero or weighing mode inoperative prices shall be shown on price-to-pay display Prices for more than one equal articles: number of articles shown on weight display without being taken for a weight price for one article shown on unit price display

OIML R 76-2: 2007 (E) Report page …./….

Requirement Testing procedures PASSED FAILED 59

Remarks supplementary display for number of articles and/or article prices 4.14.4.2 Totalization of transactions on one or several tickets: price total indicated on price-to-pay display and printed accompanied by a special word or symbol and reference to commodities whose prices are totalized if a separate ticket is issued for total all prices-to-pay shall be printed and price total shall be the algebraic sum of these printed prices Totalization of transactions from linked instruments: price-to-pay scale intervals of all connected instruments identical 4.14.4.3 Instrument used by several vendors or to serve more than one customer at the same time: connection between transactions and vendor or customer identified 4.14.4.4 Canceling previous transactions: transaction is already printed: the price-to-pay cancelled shall be printed with comment transaction not yet printed and displayed to customer: transaction clearly differentiated from normal transactions 4.14.4.5 Printing additional information: clearly correlated to transaction and does not interfere with assignment of weight value to unit symbol Price labeling instruments requirements 4.13.8, 4.14.3 (paragraphs 1 and 5), 4.14.4.1 4.16 (paragraph 1) and 4.14.4.5 are met Display: for weight possibility to verify values of unit price and preset tare during the use of the instrument Printing: prevention of printing below Min labels with fixed values of weight, unit price and price-to-pay allowed provided weighing mode is inoperative Mobile instruments used outside means to indicate that the limiting value of tilting has been 4.18.1 exceeded and to inhibit printout and data transmission automatic zero-setting or tare balancing operation after each moving of the vehicle Weighing window Existent Non-existent indication when instrument is not in the weighing window and the printout and data transmission is inhibited equipped with an appropriate protection system if the load measuring device is sensitive to moving or driving influences prevention of wrong weighing results if the cardanic suspension system or load receptor comes into contact with the surrounding frame construction 4.18.2 Other mobile instruments not to be used outside with a leveling device and a level indicator the leveling device shall be operated easily without tools appropriate inscription pointing the user to the necessity of leveling after each movement

OIML R 76-2: 2007 (E) Report page …./….

17.3 Electronic weighing instruments Requirement Testing procedures PASSED FAILED Remarks Disturbances indication of significant faults in the display does not lead to 5.1.1 confusion with other messages 5.2 Acting upon significant faults in case 5.1.1 b): instrument made automatically inoperative1, or visual or audible indication until user takes action or fault disappears1 Display check 5.3.1 Upon switch-on: signs of indication are active and non-active long enough to be checked by operator External equipment 5.3.6 Interfaces (mechanical, electrical, logical) do not allow: 󰂃 functions and measurement data to be inadmissibly influenced by peripheral devices, or other connected instruments, or disturbances 󰂃 displaying data which could be mistaken for a weighing 5.3.6.1 result 󰂃 falsifying weighing results (displayed, processed, stored) 󰂃 changing adjustment factor or adjusting the instrument (except authorized cases) 󰂃 falsifying displayed primary indications (direct sales) interfaces that do not fulfill 5.3.6.1 can be secured 5.3.6.2 interfaces transmit data so that peripheral device can meet 5.3.6.3 requirements metrologically relevant functions performed or initiated 5.3.6 through the interface meet relevant requirements of R 76-1 1 Checked by verifying the compliance with documents or by simulating faults; this check does not duplicate the disturbance tests 12.1-12.7. 60

OIML R 76-2: 2007 (E) Report page …./….

17.4 Software-controlled digital devices and instruments Requirement Testing procedures PASSED FAILED Remarks Devices with embedded software Existent Non-existent declaration of the manufacturer that the software: 5.5.1 G.1 󰂃 is used in a fixed hardware and software environment, and 󰂃 cannot be modified or uploaded by any means after securing/ verification the software documentation contains: 󰂃 description of the legally relevant functions 󰂃 description of the securing means (evidence of an intervention) 󰂃 software identification 󰂃 description how to check the actual software identification the software identification is: 󰂃 clearly assigned to the legally relevant software and functions 󰂃 provided by the instrument as documented Personal computers, instruments with PC components, and other instruments, devices, modules, and elements with programmable or loadable legally relevant software Existent Non-existent the legally relevant software is: 5.5.2.2 d G.2.1 󰂃 documented with all relevant information 󰂃 protected against accidental or intentional changes 5.5.2.2 a evidence of intervention is available until the next 5.5.2.2 a verification / inspection 5.5.2.2 G.2.2.1 Operation system / programs not accessible for the user description of all commands via keys or interfaces declaration of completeness of commands 5.5.2.2 G.2.2.2 Operating system / programs accessible for the user checksum or signature generated over the machine code of the legally relevant software legally relevant software cannot be started if the code is falsified G.2.2.3 In addition to the cases G.2.2.1 or G.2.2.2 device-specific parameters sufficiently protected audit trail for the protection of the parameters and description some practical spot checks performed 5.5.2.2 b G.2.3 Software interfaces if there is associated software providing other than measuring functions, the legally relevant software part: 󰂃 is separated from associated software 󰂃 identified 󰂃 cannot be influenced by the associated software program modules of legally relevant software are defined and separated from the modules of associated software by a defined protective software interface protective software interface itself is part of the legally relevant software description and definition of functions of the legally relevant software that can be released via the protective software interface description and definition of parameters that may be exchanged via the protective software interface 61

OIML R 76-2: 2007 (E) Report page …./….

Requirement Testing procedures PASSED FAILED Remarks description of the functions and parameters conclusive and complete each documented function and parameter does not contradict with the requirements of this Recommendation appropriate instructions for the application programmer concerning the protectiveness of the software interface 5.5.2.2 c G.2.4 Software identification the legally relevant software is identified by a software identification the software identification: 󰂃 covers all program modules of the legally relevant software and the type-specific parameters at runtime of the instrument 󰂃 is easily provided by the instrument 󰂃 can be compared with the reference identification fixed at type approval spot checks whether the checksums (signatures) are generated and work as documented there exists an effective audit trail Data storage devices (DSD) Existent Non-existent DSD realized with embedded software (examine software acc. to G.1) Yes No 5.5.3 G.3.1 DSD realized with programmable/loadable software (examine software acc. to G.1) Yes No documentation with all relevant information sufficient storage capacity for the intended purpose 5.5.3.1 G.3.2 data are stored and given back correctly sufficient description of measures to prevent data loss storage of all relevant information necessary to reconstruct an 5.5.3.2 G.3.3 earlier weighing, i.e. gross, net, tare values, decimal signs, units, identifications of the data set, instrument number, load receptor, (if applicable), checksum / signature of the data set stored protection of the stored legally relevant data against 5.5.3.3 G.3.4 accidental or intentional changes protection of the stored legally relevant data at least with a parity check during transmission to the storage device protection of the stored legally relevant data at least with a parity check of a storage device with embedded software (5.5.1) protection of the stored legally relevant data by an adequate checksum or of a storage device with programmable or loadable software (5.5.2) identification and indication of the stored legally relevant 5.5.3.4 G.3.5 data with an identification number record of the identification number on the official transaction medium, i.e. on the print-out automatic storage of the legally relevant data 5.5.3.5 G.3.6 a device subject to legal control prints or displays the stored 5.5.3.6 G.3.7 legally relevant data for verifying 62

因篇幅问题不能全部显示,请点此查看更多更全内容

Copyright © 2019- baomayou.com 版权所有 赣ICP备2024042794号-6

违法及侵权请联系:TEL:199 18 7713 E-MAIL:2724546146@qq.com

本站由北京市万商天勤律师事务所王兴未律师提供法律服务